Vitalograph In2itive Spiromter |
Subjects
|
Automatic population of database from Spirotrac. Name, ID; Age, Height, Gender; Smoking status, Body Mass Index and more depending on model. Storage of up to 10,000 subjects
|
Test Types
|
Single breath tests, flow/volume loops, multi-breath testing, tidal breathing and combined VC/FVC type test methods supported |
Flow Technology
|
Fleisch Pneumotachograph (No3 size) |
Resolution
|
10ml volume; 0.01 L/s flow
|
Data Storage
|
Up to 10,000 subjects and 100's of test sessions per subject
|
Accuracy when in Operating Range
|
Volumes |
Volumes: better than ± 3% (Max 10L / Min 0L)
|
Flows |
Flows: better than ± 10% (Max 16L/s Min 0.02L/s) |
Linearity |
Linearity: ± 5% in range 0.1L/s to 16L/s
|
Power Supply |
PowerSAFE™: Input 100 - 240V AC 50-60Hz, output 5V DC |
Battery Pack
|
Rechargeable Lithium Polymer 3.7v 2000mAH
|
Dimensions
|
Device: 160 x 100 x 45mm (with flowhead attached)
|
Net Weight
|
Device: 230 g
|
Gross weight and Size (Packed)
|
Weight: 1.5kg, Packed Size: 213 x 199 x 158mm |
Connectivity |
USB 2.0 interface |
Max. Test Duration |
20s FVC; 30s VC |
Suggested Interpretation |
User selectable |
Selectable Parameters (model dependent) |
VC; IVC; IC; VT (TV); TLC; RV; IRV; ERV; FRC; FVC; FIVC; FEV1; FEV3; FEV6; FVC; FEV1/VC; FEV1/FVC; FEV3/VC; FEV3/FVC; FEV1/FEV6; FEF75; FEF50; FEF25; FEF25-75; FEF25-75/FVC; FIV1; PIF, FIV1/FIVC, FIF25, FIF50, FIF75, FEF50/FIF50, FET, MVVind, FEV1 Ratio; FEV0.5; PEF L/min; PEF L/s; FEF 0.2-1.2; FEF 75-85%; FEF25%; FEF50%; FEF75%; FMFT; FIF25%; FIF50%; FIF75%; PIF; FEV0.75; Lung Age and more |
Predicted Values Selectable (model dependent) |
ERS/ECCS; NHANES, Polgar, Pereira, Berglund, Forsche, Gutierrez, Hedenström, Taiwan; Knudson, Crapo; Hsu; KNLW; Viljanen; SEPAR; Gulsvik; SBPT; Tamura; Ip; Quanjer; Wang; Dockery and more |
Performance Standards |
ISO 16782:2009, ISO 23747:2007, ATS/ERS 2005 |
Safety Standards |
IEC 60601-1:2005 |
Medical Safety Standards |
Medical Devices Directive 2007/47/EC |